Tab 4 4.4 Adjusting the cephalometer (XG
Plus
/ XG 5, if ceph is installed)
4.
4
– A horizontal bar must be visible in the center of the image (A). If
this bar is visible, the exposure is OK and ...
– the two beams imaged are within the tolerance band of
± 10 mm (A).
NOTE
i
If the above criteria are not fulfilled
(B)
, the ceph main X-ray
beam direction must be adjusted.
Making the unit ready for exposure
NOTE
i
The default value for S1 was automatically determined by SIDEXIS based on
the exposure and entered in the text box of the menu.
For manual adjustment, the value displayed at this position in the text box of
the menu can be overwritten (see page 4-89).
First continue with the automatic adjustment. Manual determination of the ad-
justment value is required only if you fail to reach your goal via automatic ad-
justment (see page 4-88).
5.
Make SIDEXIS XG ready for exposure:
Click
IMAGE
ACQUISITION
SIDEXIS screen
The exposure dialog box showing the exposure status
appears in SIDEXIS.
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...