båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
5 – 101
Tab 5
5.13 Service routine S017
5.13
S017.15: Activating/deactivating the acoustic signal for end of exposure
5.13.13
S017: Test step 15
z
Start test step 15 in the same way as described on page 5-78 or by
pressing the
Service key
in test step 14.
NOTE
i
If you go to test step 15 from test step 14 by pressing the Service key, the
security access confirmation will be skipped.
Activating/deactivating the acoustic
signal for end of exposure
1.
Select the code for activating or deactivating the acoustic signal using the
arrow keys
of selection field 1.
00 = acoustic signal for end of exposure = OFF
01 = acoustic signal for end of exposure = ON
2.
After the code is selected the
Memory key
lights up. To save the
changes, first press the
Memory key
(R key lights up) and then press
the
R key
.
1
2
3
+
-
00
XG
Plus
:
Easypad touchscreen
(
n
R
15
n
(
1
2
3
______00________
XG 5 / 3 / 3 PPE:
Multipad
R
1
2
3
+
-
01
XG
Plus
:
Easypad touchscreen
R
)
n
)
o
15
S17
(
n
R
n
(
1
2
3
______01________
XG 5 / 3 / 3 PPE:
Multipad
(
o
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...