5.14 Service routine S018
5.14
S018.4: Check of the height adjustment sensor system
5.14.3
S018: Test step 4
Selecting test step 4
1.
Select test step 4 in selection field 2 with the
arrow keys
and confirm your
selection by pressing the
Service key
.
The current height position is displayed in selection field 1.
NOTE
i
Following the selection of the service routine, the stand can be
moved up or down up to the limit switches using the UP/DOWN keys
on the control console. The "soft limit positions" set by the software are ig-
nored in this case.
The first three patient symbol keys (from the left) indicate the switching state
of the limit switches:
First patient symbol key = State of correction switch
Second patient symbol key = State of bottom limit switch
Third patient symbol key = State of top limit switch
1
2
3
+
-
S018
4
n
(
1
2
3
S018_______4____
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE:
Multipad
+
-
(
n
(
n
(
o
(
o
1
2
3
1.342mm
n
(
1
2
3
_1342_
MM
________
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE:
Multipad
S18
4
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...