båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
6 – 79
Tab 6
6.20 Replacing circuit boards
6.20
NOTE
i
In systems that already run with a system software version V02.27 or higher,
please check that there is a XML file in the PDATA/P2K_Config with the net-
work name of the system (Extended details). This file contains up-to-date in-
formation about the previous system configuration! See also section 1.10, on
page 1-32).
16.
Start service routine
S034.7
and select
ring type 02
(see page 5-145).
17.
Start service routine
S017
and perform the system configuration (test
step 2-23) (see page 5-78).
NOTE
i
For systems with hardware version< BA (until October 2006):
Circuit board DX41 must be configured with service routine S017.9.
NOTE
i
For the XG 3: Service routine S017.17 must be set according to the serial
number of the unit (see page 5-103).
18.
If the minimum travel height of the system must be limited:
Set the travel height with service routine
S018.2
. (see page 5-110).
19.
Perform another software update to the current system software version
as described in section 1.9.3. This updates all modules in accordance
with the configuration.
20.
Perform the complete system adjustment (see chapter 4):
– PAN adjustment
– CEPH adjustment, if installed
21.
Open the "Extended Details“ via SIXABCON.
This generates an XML file (with the system parameters) which is saved
under the network name of the system in the PDATA/P2K_Config folder
(see also section 1.10 on page 1-32).
z
The process is completed.
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...