6.19 Replacing the light barriers
Tabs 6
The following light barriers can be replaced:
In the panoramic X-ray unit
z
Light barrier at actuator 1, ON position:
V1_1
z
Light barrier at actuator 2, ON position:
V1_2
z
Light barrier at ring motor, starting position of rotation:
V1_3
z
Light barrier at HA motor, height adjustment:
V1_4
In the cephalometer
z
Light barrier for patient diaphragm position detection, ceph:
V91_1
z
Light barrier for sensor position detection, ceph:
V91_2
ATTENTION: What to do after replacement?
After replacing V1_1, 2 and 3:
– Perform a complete unit adjustment (see chapter 4).
After replacing V1_4:
– Perform a function check.
After replacing V91_1 and 2:
– Execute S034.4 and S034.5 (see pages 5.19.1 ff.)
– Check the CEPH adjustment by taking diaphragm test exposures:Fixed
point of rotation and Quickshot
– Perform a CEPH unit adjustment if necessary (see chapter 4).
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...