4.3 Adjusting the panoramic X-ray unit
4.
3
Making the unit ready for exposure
NOTE
i
The default values for Sx and Sy were automatically determined by SIDEXIS
based on the exposure and entered in the text boxes of the menu.
For manual adjustment, the values displayed at this position in the text boxes
of the menu can be overwritten (see page 4-31).
First continue with the automatic adjustment. Manual determination of the ad-
justment values is required only if you fail to reach your goal via automatic ad-
justment (see page 4-30).
5.
Make SIDEXIS XG ready for exposure:
Click
IMAGE
ACQUISITION
SIDEXIS screen
The exposure dialog box showing the exposure status
appears in SIDEXIS.
XG
Plus
:
Display on the Easypad
The Easypad touchscreen shows service routine
S030.2
as well as the related exposure parameters.
XG 5:
Display on the Multipad
The single-line display of the Multipad shows the exposure parameters
of service routine
S030.2
.
1
2
3
+
-
60 kV/3 mA
0.20 s
S30
2
1
2
3
60/3____0.20____
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...