4.3 Adjusting the panoramic X-ray unit
Tabs 4
Manual adjustment of the PAN filter
The manual adjustment procedure is similar to the one for automatic adjust-
ment. The only difference is that the default adjustment value automatically
determined by SIDEXIS is overwritten by a manually determined adjustment
value in the
PAN
-
FILTER
submenu.
Determining the adjustment value for
Fx manually
z
Measure distance
Fx
with the SIDEXIS measuring ruler.
NOTE
i
Tip:
To facilitate the measuring procedure, you can color the image in SIDEX-
IS (see also SIDEXIS Operator’s Manual).
Overwriting the default value for Fx
z
Replace the default value for
Fx
by entering the measured value in the text
box of the
PAN
-
FILTER
submenu.
NOTE
i
For information on the direction of displacement (input of +/- sign in the menu)
see page 4-7. Use points as decimal separators!
z
Proceed with step 5 of the adjustment procedure.
Fx
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...