4.5 TSA sensor adjustment (XG
Plus
full version only)
4.
5
Evaluating the X-ray image
7.
Evaluate the X-ray image:
– The exposed diaphragm area must lie horizontally and vertically
centered in the image field (A).
– A white border surrounding the image on all sides must be visible.
The maximum density must lie in the center of the diaphragm area
(A).
NOTE
i
If these criteria are not yet fulfilled, repeat the adjustment procedure starting
with step 5.
NOTE
i
If you do not reach your goal via automatic adjustment, repeat the adjustment
procedure with manually determined adjustment values (see page 4-113).
Adjustment: ok