6.20 Replacing circuit boards
Tabs 6
Case K: X-ray tube assembly from
external system,
ORTHOPHOS
XG 5 / 3 (not XG 3 PPE)
System software status
V02.20 or higher.
CAUTION
– The replacement of an X-ray tube assembly from an external device is
possible if the X-ray tube assembly and the system have at least the
software version V02.20.
– The replacement is only possible within the same system class, i.e.
X-ray tube assembly must come from an XG 5 or XG 3
-
system if it is to
be installed in an XG 5 or XG 3 system
1.
Switch the unit ON.
CAUTION
Do not acknowledge any error messages displayed at this point!
2.
Install the current SIDEXIS software version (V01.53 or higher).
NOTE
i
If the current SIDEXIS version is a patch version, the previous officially re-
leased main version of SIDEXIS has to be installed before the current version
can be installed.
3.
Perform a software update to Version V02.25 or higher (automatic up-
date) (see also section 1.9.5).
4.
Switch the unit OFF.
Wait approximately 1 minute. Then switch the unit ON again.
Error message
E6 15 05
(undefined system serial number) is displayed.
5.
Acknowledge the error message with the
R key
.
6.
Open the Service menu (see page 5-10).
7.
Start service routine
S008.3
(see page 5-50).
8.
Get the unit serial number from the rating plate of the unit and enter it (see
page 5-50).
NOTE
i
Any serial number which is unknown to the unit will not be accepted by the
unit. The serial number entered must be identical with the one on the rating
plate of the unit. If an inadmissible serial number is entered, the input will not
be accepted and the serial number can be entered again.
9.
Switch the unit OFF.
Wait approximately 1 minute. Then switch the unit ON again.
10.
Install the current SIDEXIS software version (V01.53 or higher).
NOTE
i
If the current SIDEXIS version is a patch version, the previous officially
released main version of SIDEXIS has to be installed before the current
version can be installed.
11.
Perform a software update to Version V02.27 or higher (automatic up-
date) (see also section 1.9.5).
E6 15 05
(
R
R
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...