59 38 399 D3352
4 – 108
D3352.076.01.13.02
07.2008
4.5 TSA sensor adjustment (XG
Plus
full version only)
Tabs 4
SIDEXIS screen
The exposure dialog box showing the exposure status
appears in SIDEXIS.
XG
Plus
:
Display on the Easypad
The Easypad touchscreen shows service routine
S030.4
as well as the related exposure parameters.
Starting the exposure
3.
Take an exposure (60 kV/3 mA; 0.20 s):
– Press the
R key
to move the unit back to the starting position.
The procedure is completed when the status message “ORTHOPHOS is
ready for exposure“ appears on the touchscreen and the exposure
parameters of service routine
S030.4
are displayed.
– Press the release button. Hold down the release button until image ac-
quisition is completed and the preview image appears in the exposure
dialog box.
1
2
3
+
-
60 kV/3 mA
0.20 s
S30
4
R
WAIT
!
UNTIL
UNIT
IS
IN
STARTING
POSITION
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...