båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
5 – 125
Tab 5
5.16 Service routine S021 (not for XG 3 / 3 PPE)
5.16
S021.3: Function test of TSA sensor adjustment
Moving the TSA sensor holder
3.
Move the TSA sensor holder by alternately pressing
patient symbol keys
a
and
d
.
When key P1 is pressed, the TSA sensor holder moves to the
PAN/CEPH position for Set key P4 to the TSA position.
When the action is completed, the result is displayed in selection field 2:
0000
=
action OK
EEEE =
action error
Quitting the service routine
XG
Plus
:
Pressing the
Service key
or the
double arrow key
returns you to the
service routine selection menu.
To quit the Service menu and return to the main menu, press the
double arrow key
.
XG 5:
To return to the service routine selection menu, press the
up arrow key
above selection field 3.
To quit the Service menu and return to the Main menu, press the
up arrow key
above selection field 3.
T
n
(
1
2
3
1
2
3
XG
Plus
:
Easypad touchscreen
XG 5:
Multipad
3
S21
(
n
)
n
a
d
a
d
T
1
2
3
0000
n
(
1
2
3
_______0000____
XG
Plus
:
Easypad touchscreen
XG 5:
Multipad
S21
3
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...