6.20 Replacing circuit boards
6.20
Case E: DX11 from an external device,
ORTHOPHOS
XG 5 / 3 (not XG 3 PPE)
System software status
V02.20 or higher.
CAUTION
– The replacement of a DX11 from an external device is possible if
the DX11 board and the system have at least the software version
V02.20.
– The replacement is only possible within the same system class, i.e.
DX11 must come from an XG 5 or XG 3 device if it is to be installed in
an XG 5 or XG 3 system.
CAUTION
After inserting the board, you must reconfigure the IP address to match the IP
address of the existing X-ray component. Before you set the unit to the correct
IP address, make sure that this address has not been assigned to any other
device.
1.
Switch the unit ON.
CAUTION
Do not acknowledge any error messages displayed at this point!
2.
Perform a software update to a software version V02.25, however, ideally
to the latest software version via the automatic update option (see SIRO-
NA internet website, Dealer domain (under Product Info
Æ
X-ray Systems)
(see also Section 1.9.6).
NOTE
i
The current update file can be found on the ORTHOPHOS XG Software CD.
This CD is supplied along with every spare DX11 PC board and is also inclu-
ded in the language-specific set. The contents of this CD can also be down-
loaded (under Product Info
Æ
X-ray Systems) from the Dealer domain of the
SIRONA internet website at www.sirona.com
CAUTION
If the DX11 used already has the same software version as the overall unit, a
software update to this version must nevertheless be performed so that an ad-
ministrative entry can be made in the memory of the DX11.
3.
Switch the unit OFF.
Wait approximately 1 minute. Then switch the unit ON again.
Error message
E6 15 05
(undefined system serial number) is displayed.
4.
Acknowledge the error message with the
R key
.
5.
Open the Service menu (see page 5-10).
6.
Start service routine
S008.3
(see page 5-50).
7.
Get the unit serial number from the rating plate of the unit and enter it (see
page 5-50).
NOTE
i
Any serial number which is unknown to the unit will not be accepted by the
unit. The serial number entered must be identical with the one on the rating
plate of the unit. If an inadmissible serial number is entered, the input will not
be accepted and the serial number can be entered again.
8.
Switch the unit OFF.
Wait approximately 1 minute. Then switch the unit ON again.
E6 15 05
(
R
R
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...