![background image](http://html1.mh-extra.com/html/sirona/orthophos-xg-3-ds/orthophos-xg-3-ds_service-manual_1285016464.webp)
59 38 399 D3352
6 – 32
D3352.076.01.13.02
07.2008
6.10 Removing the light localizers from the ring (for the XG
Plus
and XG 5, if present)
Tabs 6
6.10
Removing the light localizers from the ring
(for the XG
Plus
and XG 5, if present)
In XG
Plus
and XG 5 units with an older serial number, a light localizer is
located on the left and on the right bottom side of the ring. Both of these light
localizers must be removed for TSA operation. To do this, proceed as follows:
1.
Unplug the cables of the light localizers from connectors X405 and X406
on board DX61 and detach the cables from the cable clamps.
2.
Unscrew the holders including the light localizers from the ring.
NOTE
i
Depending on the serial number of the unit, different versions of light localizers
may be installed.
CAUTION
For "TSA ready" units, make sure that cable L20 (TSA motor) and cables L21
(light barriers) are once again correctly installed in the cable clamps (see page
6-45). Tighten the cables slightly if necessary.
X406 X405
Light localizer
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...