59 38 399 D3352
2 – 46
D3352.076.01.13.02
07.2008
2.5 List of error messages
Tab 2
*)
TTP = Time Trigger Protocol
E1
61
60
Voltage error on DX61,
40 V or 28 V
Note:
If several drives are operated simulta-
neously under unfavorable circum-
stances, this may cause voltages to
fall below their minimum tolerances.
z
Acknowledge error and repeat procedure
If the error occurs again...
z
Check height adjustment (soft start),
replace DX1 if necessary
z
Check voltages at X501 on board DX61 accord-
ing to wiring diagrams,
replace diaphragm unit if necessary
z
Check cable L11,
replace if necessary
z
Check voltage path from DX32 to DX1 (X102)
according to wiring diagrams,
replace component if necessary
E6
61
70
Error during the configuration of the
diaphragm unit
If this error occurs after board DX11 has been
replaced:
z
Check configuration of diaphragm unit via
service routine S017.17
Configure correctly if necessary
If this error occurs after the diaphragm unit has
been replaced:
z
Check order number of inserted diaphragm unit
Replace diaphragm unit if necessary
6-34,
Error code
Description
Actions required
see page
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...