1.5 Operation notes
1.5
1.5
Operation notes
Rated line voltage
The ORTHOPHOS XG X-ray unit can be operated in the following
rated line voltage ranges:
z
200 V
-
240 V
z
50/60 Hz
The permissible line voltage fluctuations are as follows:
z
200 - 230 V: ± 10 %
The internal line impedance must not exceed max. 0.8
Ω
.
NOTE
i
Only permanent electrical connection of the system is allowed in Germany.
Remote control
The system can be equipped with...
z
a 1 - 3 m coiled cable with release button inside the treatment room or ...
z
a remote control with or without coiled cable located outside the X-ray
room (see installation instructions).
Warm-up time
After it is switched ON, the system requires a warm-up time of approx. 1 min.
Self-adjustment routine
At the same time, a mechanical and electronic self-adjustment routine is exe-
cuted. If a button is pressed during the self-adjustment routine, an error mes-
sage is displayed on the Multipad (XG 5 / 3 / 3 PPE) or Easypad (XG
Plus
).
Cooling period
The cooling period between two exposures is maintained by an automatic
exposure blocking function according to the pulse/pause ratio. A countdown
of the waiting time is displayed on the Multipad (XG 5 / 3 / 3 PPE) or
Easypad (XG
Plus
).
Turn-off time
The turn-off time must amount to at least 60s.
Demo units
If the X-ray unit is to be presented as a demo unit at trade fairs or exhibitions,
it must be ensured that radiation release is blocked (see “Demo mode –
Operation without radiation release” on page 1-13).
Software version
The overall system software version is determined by the software statuses of
the EEPROMs on the boards (see “List of software versions” on page 1-16).
Wireless phone interference with
medical electrical equipment
To ensure safe operation of medical electrical equipment, the use of mobile
wireless phones in practice or hospital environments is prohibited.
Disposal
The X-ray tube assembly contains a tube with potential implosion hazard,
a small amount of beryllium, a lead lining as well as mineral oil.
Error messages
Error messages are displayed on the single-line display of the Multipad
(XG 5 / 3 / 3 PPE) or on the touchscreen of the Easypad (XG
Plus
).
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...