59 38 399 D3352
4 – 30
D3352.076.01.13.02
07.2008
4.3 Adjusting the panoramic X-ray unit
Tabs 4
Manual adjustment of the PAN diaphragm
The manual adjustment procedure is similar to the one for automatic adjust-
ment. The only difference is that the default adjustment values automatically
determined by SIDEXIS are overwritten by manually determined adjustment
values in the
PAN
-
DIAPHRAGM
submenu.
Determining adjustment values for Sx
and Sy manually
z
Measure distances
Sx
and
Sy
with the SIDEXIS measuring ruler.
NOTE
i
Tip:
To facilitate the measuring procedure, you can color the image in SIDEX-
IS (see also SIDEXIS Operator’s Manual).
S2
Sy
Sx
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...