59 38 399 D3352
5 – 38
D3352.076.01.13.02
07.2008
5.6 Service routine S005
Tabs 5
S005.7: Setting continuous operation of the tube assembly fan
The operating mode of the fan is displayed in selection field 1:
01 = continuous operation of the fan
00 = controlled operation of the fan
Switching on continuous fan operation
3.
To switch the fan to continuous operation, select code 01 in selection
field 1 with the arrow keys.
The
Memory key
(XG
Plus
) or the LED above the
Memory key
(XG 5 / 3 / 3 PPE)
lights up.
1
2
3
-
+
00
7
n
(
1
2
3
00____________
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE:
Multipad
(
n
(
n
R
1
2
3
-
+
01
7
n
(
1
2
3
00____________
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE:
Multipad
(
n
(
n
R