59 38 399 D3352
4 – 100
D3352.076.01.13.02
07.2008
4.4 Adjusting the cephalometer (XG
Plus
/ XG 5, if ceph is installed) Tabs 4
4.4.7
Checking and adjusting the alignment of the
ear plugs
Checking the ear plug alignment
NOTE
i
The sensor must be plugged into the sensor slot on the cephalometer.
The sensor must not be plugged into the sensor slot on the panoramic X-ray
unit.
Preparations
XG
Plus
:
z
Select the
CEPH
mode in the main menu on the Easypad.
XG 5:
z
Select one of the ceph programs on the Multipad.
Position of the ear plug holders
1.
Move the
ear plug holders
completely apart and swing them
into the beam path
.
Fitting the adjusting caps
2.
Fit adjusting caps
D
onto the ear plugs and secure them with adhesive
tape.
z
Fit the black adjusting cap on the outside (sensor side)
z
Fit the transparent adjusting cap on the inside (tube assembly side).
D
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...