59 38 399 D3352
5 – 80
D3352.076.01.13.02
07.2008
5.13 Service routine S017
Tabs 5
S017.2: Configuring the hardware version
Checking/setting the jumper position
5.
Check the jumper settings of sockets
306 (DX61, diaphragm)(only if XG 3 / 3 PPE= 0081),
309 (DX91, ceph sensor) and
503 (DX81, ceph sensor)
on board DX1.
DX1
Socket X309
Jumper inside
Cephalometer not connected,
i.e. connector not plugged in
0001 and 0041
Socket X503
NOTE
i
For more information on the correct jumper
positions and CAN bus wiring, refer to sec-
tion 3.2.2.
Socket X309
Jumper outside
Cephalometer connected,
i.e. connector plugged in
0003, 000, 0043 and 0045
Socket X503