5 – 41
Tab 5
5.7 Service routine S007
5.
7
S007.1: Display error logging memory
5.7.1
S007: Test step 1
Selecting test step 1
1.
Select test step 1 in selection field 2 with the
arrow keys
and confirm your
selection by pressing the
Service key
.
2.
Select the desired error event (in example 66).
XG
Plus
:
Easypad touchscreen
You can browse between the different occurred error numbers with the arrow
keys in selection field 1.
The corresponding error message is displayed in selection field 2 (see
section 2.5). Selection field 3 displays the date and time of the error event.
You can set the step width for browsing between the error numbers with the
patient symbol keys.
The currently selected patient symbol key is lit. A step width of 1 is preset (left
patient symbol key is lit).
1
2
3
+
-
S007
1
n
(
1
2
3
S007_______1____
XG
Plus
:
Easypad touchscreen
XG 5/3/3 PPE:
Multipad
+
-
(
n
(
n
(
o
(
o
1
2
3
+
-
66
E6 81 07
2004-12-14; 12:31:09
(
n
1
Step width 100
Step width 10
Step width 1
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...