1.9 Software update
1.9
1.9.4
Software update for ORTHOPHOS XG 5 to
V02.30 or higher
NOTE
i
If the sensors previously used are to be used in the future as well, plug them
into the slots on the panoramic unit and/or cephalometer before you begin the
update.
1.
Switch the unit ON.
2.
Execute the software update to Version V02.30 or higher (automatic up-
date) as described on page 1-24.
3.
Switch the unit OFF.
Wait for approx. 1 minute. Then switch the unit ON again.
Error message
E6 15 04
(undefined activation data) is displayed.
4.
Acknowledge the error message with the
R key
.
5.
Perform a ceph adjustment if a cephalometer is installed (see chapter
4.4).
6.
Switch the unit OFF.
Wait for approx. 1 minute. Then switch the unit ON again.
z
The process is completed.
E6 15 04
(
R
R
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...