59 38 399 D3352
7 – 40
D3352.076.01.13.02
07.2008
7.13 Checking the protective ground wires
Tabs 7
Check whether the ground wire
resistance complies with the
specifications
z
A power source with a current of at least 0.2 A and a no-load voltage of 24
V max. and 4 V min. is required.
z
Connect the power source between the measuring points specified in the
table for at least 5 s and measure:
– the voltage drop with the voltmeter
– the current with the ammeter and
– calculate the resistance using the formula
R = U/I
Measuring point A: Central ground wire
Protective ground wire test
between...
A and
B
GNYE wire
0,1
Ω
A and
C
2. Protective
ground wire
0,1
Ω
A and
D
Housing DX32
0,2
Ω
A