59 38 399 D3352
7 – 14
D3352.076.01.13.02
07.2008
7.7 Checking the tube data
Tabs 7
1.
Remove the covers of the tube assembly.
2.
Remove jumper
A
from the MA+/MA– measuring points on board DX6.
Connect a digital ammeter to MA+ and MA– and select range 20 mADC.
3.
Reattach the cover and tighten the screw securely.
MA–
MA+
20mADC
com.
Board DX6
·
A
400VDC