5.13 Service routine S017
5.13
S017.9: Configuring operaton without DX41
5.13.7
S017: Test step 9
This service routine is used to configure operation of the system with or with-
out the module DX41. This configuration is necessary for software updates
and module replacement or board DX11 if systems with or without board
DX41 should be supported.
z
Start test step 9 in the same way as described on page 5-78 or by
pressing the
Service key
in test step 8.
NOTE
i
If you go to test step 9 from test step 8 by pressing the Service key
≤≤
, the
security access confirmation will be skipped.
Setting configuration without DX41
1.
Select the desired device status using the
arrow keys
of selection field 1.
00 = operation without DX41
(Factory setting for unit hardware versions
≥
BA)
01 = operation with DX41
(Factory setting for unit hardware versions
≤
AG)
After the device status is selected, the
Memory key
(XG
Plus
) or the LED above the
Memory key
(XG 5 / 3 / 3 PPE) lights up.
1
2
3
+
-
00
n
(
1
2
3
_______00________
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE:
Multipad
(
n
(
n
R
9
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...