4.3 Adjusting the panoramic X-ray unit
4.
3
4.3
Adjusting the panoramic X-ray unit
4.3.1
Pan needle phantom
Accessory: Pan needle phantom
In order to perform the
pan sensor adjustment
and the
symmetry adjust-
ment
you must insert the needle phantom
(A)
in the bite block holder of the
panoramic X-ray unit.
The needle phantom must be removed from the bite block holder for the
PANdiaphragm
and for the
TSA adjustment
.
CAUTION
It is essential that the needle phantom is removed from bite block holder of the
panoramic X-ray unit again before a Ceph exposure is taken; otherwise the
phantom may collide with the sensor.
NOTE
i
When fitting the needle phantom, make sure that it is
correctly oriented
. For
the adjustment of the X-ray unit, the phantom must be fitted in such a way that
the
needles point upward
.
A
Needles on
Top side
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...