3.6 Device leakage current too high
Tab 3
3.6
Device leakage current too high
Pull tube assembly cable L3 off of
connector X3 on board DX6 and per-
form measurement of leakage current
according to sect. 7.14.
• Is the leakage current OK?
Check the cable shields
(see section 7.12) and check the
cables for visible signs of damage.
• Are the cable shields and
cables OK?
yes
Replace tube assembly,
see section 6.14.
no
yes
Replace board DX32,
see section 6.1.2.
Fix cable shields and/or replace any
defective cables.
no