4 – 93
Tab 4 4.4 Adjusting the cephalometer (XG
Plus
/ XG 5, if ceph is installed)
4.
4
Evaluating the X-ray image
4.
Evaluate the X-ray image:
– The exposed diaphragm area must lie centered and straight in
the image field as well as inside the superimposed auxiliary lines
(A).
– A white border surrounding the image on all sides must be visible.
The maximum density must lie in the center of the diaphragm area
(A).
NOTE
i
If these criteria are not fulfilled
(B)
, the ceph quickshot
must be adjusted.
B
Adjustment: ok
A
Adjustment: not ok