5 – 27
Tab 5
5.5 Service routine S002
5.
5
S002.4: Radiation without rotary movement, max. radiation time selectable, with 0.8 mm Cu
5.5.3
S002: Test step 4
(not for XG 3 / 3 PPE)
Difference from S002, test step 1:
9
A filter element for a 0.8 Cu step filter is inserted in the beam path
9
The primary diaphragm is widened symmetrically to the pan
position so that the step filter still remains covered
NOTE
i
For selection and operating sequence, see S002, test step 1.
After you have selected test step 4, the primary diaphragm is moved to
the new position (opened fully and symmetrical to PAN position).
During this time, a progress indicator is displayed in selection field 1.
n
(
1
2
3
1
2
3
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE
: Multipad
4
S2