4.3 Adjusting the panoramic X-ray unit
Tabs 4
Mechanical adjustment of the fixed
diaphragm
z
Remove the covers (see also page 1-50):
– Tube assembly, front
– Tube assembly, rear
z
Loosen both screws
(A)
slightly
(approx. 1 turn).
z
Adjust the diaphragm position via eccentric screws
B
(horizontal adjust-
ment) and
C
(vertical adjustment).
–
CW rotation
of the eccentric screws:
movement of the diaphragm
to the right or upward
–
CCW rotation
of the eccentric screws:
movement of the diaphragm
to the left or downward
NOTE
i
To measure the shift, refer to the gap between the plastic support and the lead
diaphragm (D).
z
Tighten screws
A
again.
D
Plastic
support
Lead
diaphragm
D
Version delivered until approx. end of 2006
Version delivered after approx. end of 2007
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...