4 – 45
Tab 4
4.3 Adjusting the panoramic X-ray unit
4.
3
4.3.7
Adjusting the pan symmetry
(XG
Plus
/ 5 / 3 / 3 PPE)
Menu: Diaphragm/system adjustment
‡
Pan - Symmetry
Preparations
z
Insert the needle phantom in the bite block holder of the panoramic X-ray
unit (see page 4-9).
Opening
PAN
-
SYMMETRY
1.
Go to the
PAN
-
SYMMETRY
submenu.
Making the unit ready for exposure
2.
Make SIDEXIS XG ready for exposure:
Click
IMAGE
ACQUISITION