3.7 Checking the cables
3.7
3.7
Checking the cables
NOTE
i
You can use a standard Cat5 cable as a test cable for L8*, L10, L12, L40
and L37. Caution! This cable must not be permanently installed.
NOTE
i
Most cables have the same plug at both ends and are connected 1:1
(see also section "1.12 Cabling overview').
* up to unit hardware version AG
Is the cable plug connection OK?
For shielded cables:
Is the cable shield connection OK?
yes
If a substitute cable is available,
temporarily connect it and
check unit functioning.
• Is the unit functioning properly?
yes
Replace cable, see section 6.21.
Clean plug; in case of a defect,
fix plug, replace cable if necessary,
see section 6.21.
no
Repair shield,
replace cable if necessary,
see section 6.21.
no
Locate module causing fault and
replace component,
see section 6.
no
yes
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...