båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
5 – 29
Tab 5
5.6 Service routine S005
5.
6
S005.2: Automatic adjustment of preheating
5.6.1
S005: Test step 2
Selecting test step 2
1.
Select test step 2 in selection field 2 with the
arrow keys
and confirm your
selection by pressing the
Service key
.
Confirming the security access
2.
Confirm the security access by once again selecting the number of the
main routine (5) with the
arrow keys
of selection field 2 and then pressing
the
Service key
.
An inactive progress indicator in selection field 1 and the message FFFF
in selection field 2 signal that the system is ready for compensation.
(
1
2
3
+
-
S005
2
n
(
1
2
3
S005_______2____
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE
: Multipad
o
+
-
(
n
(
n
(
o
1
2
3
-
+
5
n
(
1
2
3
___________5____
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE
: Multipad
(
n
(
n
(
o
(
o
1
2
3
FFFF
n
(
1
2
3
_______FFFF____
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE
: Multipad
S5
2
E
X RAY
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...