Remote Commands
R&S
®
Spectrum Rider FPH
323
User Manual 1321.1011.02 ─ 11
Example:
CORR:TRAN2:SEL 'FSH-Z38.sectrd'
Selects the FSH-Z38 secondary transducer factor.
[SENSe:]CORRection:TRANsducer<1...2>:UNIT?
This command queries the unit (DB|DBVM|DBAM) of the current transducer factor.
The numeric suffix at TRANsducer specifies the primary or secondary transducer.
Example:
CORR:TRAN2:UNIT?
-> DBVM
Return unit (DBVM) of the secondary transducer.
Usage:
Query only
UNIT<1...2>:POWer
<Unit>
This command selects the unit of the vertical axis.
The availability of units depends on the operating mode and type of measurement.
The numeric suffix at UNIT is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<Unit>
DBM | DBMV | DBUV | DUVE | DMVE | VEMF | DBUA | VOLT |
WATT | DUVM | DUAM | V | W | V_M | W_M2 | DB | RHO |
MRHO
Note that the availability of units depends on the operating
mode.
*RST:
DBM
Example:
UNIT:POW DBM
Sets the power unit to dBm.
13.6.2.3
Configuring the Tracking Generator
The following commands configure the characteristics of the tracking generator of the
R&S Spectrum Rider.
List of commands
............................................................................................... 324
................................................................................................324
......................................................................................324
................................................................................... 325
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH