Spectrum Analyzer Mode
R&S
®
Spectrum Rider FPH
177
User Manual 1321.1011.02 ─ 11
1. Press [SPAN] key.
2. Select "Full Span" softkey.
The R&S Spectrum Rider displays the spectrum over its entire frequency range.
3. Select "Last Span" softkey.
The R&S Spectrum Rider restores the span that you have set just before display-
ing the entire frequency range.
Time domain measurements
You can also activate time domain measurements without having to enter the value
manually. When measuring in the time domain, the span is 0 Hz. In that state, the R&S
Spectrum Rider measures the signal at the current center frequency only. Instead of
displaying the spectrum, the R&S Spectrum Rider shows the signal power over a cer-
tain time period. The horizontal axis becomes the time axis. The display always starts
at 0 s and stops after the currently set sweep time.
1. Press [SPAN] key.
2. Select "Zero Span" softkey.
The R&S Spectrum Rider sets a span of 0 Hz and performs the measurement in
the time domain.
6.2.3
Configuring the Vertical Axis
All relevant settings to configure the vertical axis are available in the amplitude menu.
You can access it via the [AMPT] key.
..................................................................................177
........................................................................................178
..................................................................................... 179
....................................................................................179
....................................................................................179
...........................................................................................181
................................................................................. 181
.....................................................................................181
6.2.3.1
Setting the Reference Level
The reference level is represented graphically by the grid line at the top of the diagram.
The reference level sets the input signal gain up to the display stage. If the reference
level is low, the gain is high. That means that even weak signals are displayed clearly.
If you are measuring strong signals, you have to set a high reference level to prevent
an overload of the signal path of the analyzer and to keep the signal within the display
range. If you are measuring the spectrum of a composite signal, make sure that the
reference level is high enough to cover all signals and that all signals are within the
measurement diagram.