Remote Commands
R&S
®
Spectrum Rider FPH
444
User Manual 1321.1011.02 ─ 11
*RST:
POLYgon
Example:
DISP:TRAC:STYL LIN
Selects the display of a line trace.
FORMat:BORDer
<TransferOrder>
This command selects the format of binary data.
Parameters:
<TransferOrder>
NORMal | SWAPped
NORMal
The most significant byte is transferred first big endian).
SWAPped
The least significant byte is transferred first (little endian).
*RST:
SWAPped
Example:
FORM:BORD NORM
Changes the byte order to normal mode.
FORMat[:DATA]
<Format>[, <>]
This command selects the data format that is used for transmission of trace data from
the R&S Spectrum Rider to the controlling computer.
Note that the command has no effect for data that you send to the R&S Spectrum
Rider. The R&S Spectrum Rider automatically recognizes the data it receives, regard-
less of the format.
Parameters:
<Format>
ASCii | REAL
ASCii
Returns the data in ASCII format, separated by commas.
REAL
Returns the data as 32-bit IEEE 754 floating point numbers in
the "definite length block format".
*RST:
ASCii
In REAL, 32 format, a string of return values would look like:
#42424<value 1><value 2>...<value n>
with
#4 - representing the number of digits of the following number of
data bytes (= 4 in this example);
2524 - representing the number of following data bytes (2524,
corresponds to the 711 sweep points of the R&S Spectrum
Rider.
<value> - representing 4-byte floating point value.
Example:
FORM ASC
Selects the ASCii data format.
Remote Control - Commands
www.allice.de