![Rohde & Schwarz 1321.0615.02 User Manual Download Page 484](http://html1.mh-extra.com/html/rohde-and-schwarz/1321-0615-02/1321-0615-02_user-manual_1477771484.webp)
Remote Commands
R&S
®
Spectrum Rider FPH
477
User Manual 1321.1011.02 ─ 11
Parameters:
<Volume>
Range:
0 to 1
*RST:
0.3
Example:
SYST:BEEP:VOL 0.50
Sets the volume of the beeper to 50%.
SYSTem:BEEPer:KEY:VOLume
<Volume>
This command sets the volume of the keyboard click noise.
Parameters:
<Volume>
Range:
0 to 1
*RST:
0.3
Example:
SYST:BEEP:KEY:VOL 0.10
Sets of keyboard clicking volume to 10%.
13.6.11.5
Setting up a Network Connection
This chapter describes all commands that are used if the R&S Spectrum Rider is part
of a network.
List of commands
SYSTem:COMMunicate:LAN:ETHernet
............................................................................ 477
SYSTem:COMMunicate:LAN:GATeway
............................................................................ 477
SYSTem:COMMunicate:LAN:SUBMask
............................................................................477
SYSTem:COMMunicate:SOCKet:ADDRess
...................................................................... 478
SYSTem:COMMunicate:SOCKet:DHCP[:STATe]
............................................................... 478
SYSTem:COMMunicate:SOCKet:PORT
........................................................................... 478
SYSTem:COMMunicate:LAN:ETHernet
This command queries the MAC address of the R&S Spectrum Rider.
This command is a query and therefore has no *RST value.
Example:
SYST:COMM:LAN:ETH?
Returns the MAC address.
SYSTem:COMMunicate:LAN:GATeway
<Gateway>
This command sets the gateway in the LAN.
Parameters:
<Gateway>
String containing the identifier of the gateway.
SYSTem:COMMunicate:LAN:SUBMask
<Submask>
This command sets the subnet mask of the R&S Spectrum Rider
Remote Control - Commands