CALCulate<1...2>:MARKer<1...6>:MINimum[:PEAK]
This command positions a marker on the current trace minimum.
If necessary, the corresponding marker is activated first.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Example:
CALC:MARK2:MIN
Moves marker 2 to the trace minimum.
Usage:
Event
CALCulate<1...2>:MARKer<1...6>:X
<FrequencyOrTime>
This command positions a marker on a particular coordinate on the horizontal axis.
If one or more delta markers (2 to 6) are already active, the command turns these delta
markers into normal markers.
Note that it is possible to place the marker outside the visible trace. In that case, this
value is invalid.
If necessary, the corresponding delta marker is activated first.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Parameters:
<FrequencyOrTime> Indicates the coordinate on the horizontal axis.
The unit in spectrum analyzer mode depends on the measure-
ment, e.g. Hz for measurements in the frequency domain and
seconds for measurements in the time domain.
Range:
Maximum span.
Default unit: GHz
Example:
CALC:MARK2:X 10.7MHz
Positions marker 2 to frequency 10.7 MHz.
CALCulate<1...2>:MARKer<1...6>:X:SLIMits<1...2>[:STATe]
<State>
This command turns marker search limits on and off.
The search limit restricts the evaluation range of the trace when "set marker" functions
are performed. For example, marker set to peak with command
on page 343 sets the marker only at the peak of
the trace within the search limit.
The numeric suffix at CALCulate is irrelevant for this command.
The numeric suffix at MARKer selects the marker. The numeric suffix at SLIMits selects
the limit line.