If the VBW is automatically calculated by the R&S Spectrum Rider with the "Auto
VBW" function, the VBW is calculated as follows:
–
VBW
≥
3 x RBW
The R&S Spectrum Rider then selects the smallest possible VBW with regard to
the available step size.
●
Selecting the detector:
It is best to use the RMS detector. This detector correctly indicates the power irre-
spective of the characteristics of the signal to be measured. The whole IF envelope
is used to calculate the power for each measurement point. The IF envelope is
digitized using a sampling frequency which is at least five times the resolution
bandwidth which has been selected. Based on the sample values, the power is cal-
culated for each measurement point using the following formula:
with
S
i
= linear digitized video voltage at the output of the A/D converter
N = number of A/D converter values per measurement point
P
RMS
= power represented by a measurement point
When the power has been calculated, the power units are converted into decibels
and the value is displayed as a measurement point.
In principle, the sample detector would be possible as well. Due to the limited num-
ber of measurement points used to calculate the power in the channel, the sample
detector would yield less stable results.
6.1.6.1
Selecting a Standard
For more information on selecting a standard, see
Chapter 6.1.3.1, "Selecting a Stan-
6.1.6.2
Configuring the Measurement
In addition to creating and editing standards with the R&S InstrumentView software,
the R&S Spectrum Rider also provides functionality to set up a measurement configu-
ration.
Setting the number of transmission channels
1. Press [MEAS] key.
2. Select "Channel Settings" softkey.
The R&S Spectrum Rider opens a submenu to select the channel settings.
3. Select the "Tx Channels" menu item.
The R&S Spectrum Rider opens an input field to define the number of transmission
channels.