![Rohde & Schwarz 1321.0615.02 User Manual Download Page 297](http://html1.mh-extra.com/html/rohde-and-schwarz/1321-0615-02/1321-0615-02_user-manual_1477771297.webp)
Receiver Mode
R&S
®
Spectrum Rider FPH
290
User Manual 1321.1011.02 ─ 11
Selecting the trace mode and working with memory traces
For more information on trace modes and memory traces, see
on page 190 and
Chapter 6.2.6.4, "Working with Memory
12.2.2
Selecting the Measurement Bandwidths for EMI Measurements
The receiver mode adds 6 dB resolution bandwidths to the 3 dB resolution bandwidths
already available in other operating modes. The 6 dB bandwidths are special band-
widths that are necessary for measurements according to CISPR16.
If automatic selection of the CISPR bandwidth is on, the R&S Spectrum Rider selects
an appropriate CISPR bandwidth, depending on the receive frequency.
●
Frequencies from 9 kHz to 150 kHz: 200 Hz CISPR bandwidth
●
Frequencies from 150 kHz to 30 MHz: 9 kHz CISPR bandwidth
●
Frequencies from 30 MHz to 1 GHz: 120 kHz CISPR bandwidth
You can also select a 3 dB or 6 dB bandwidth manually.
1. Press [BW] key.
2. Select "Manual RBW" softkey to select a 3 dB bandwidth.
3. Alternatively, select "Manual CISPR BW" softkey to select a 6 dB bandwidth.
The R&S Spectrum Rider opens an input field to define the CISPR bandwidth.
4. Select the bandwidth with the rotary knob or by entering the corresponding num-
ber.
5. Select "Auto CISPR BW" to select the 6 dB bandwidth automatically according to
the list above.
12.2.3
Defining the Measurement Time
The measurement time is the time that the R&S Spectrum Rider collects data at each
measurement frequency to calculate the results for that frequency according to the
detector you have selected.
You can define a measurement time between 5 ms and 1000 s.
1. Press [SWEEP] key.
2. Select "Meas Time" softkey.
The R&S Spectrum Rider opens an input field to define the measurement time.
3. Define the measurement time.
Configuring Measurements in Receiver Mode
www.allice.de
Allice Messtechnik GmbH