[SENSe:]CORRection:TRANsducer<1...2>:SELect
<TransducerName>
This command selects a transducer factor.
If <name> does not exist yet, a new transducer factor is created.
The numeric suffix at TRANsducer specifies the primary or secondary transducer.
Parameters:
<TransducerName> String containing the filename of the transducer factor.
If the file does not exist, the R&S Spectrum Rider creates a new
transducer factor.
Example:
CORR:TRAN2:SEL 'FSH-Z38.sectrd'
Selects the FSH-Z38 secondary transducer factor.
[SENSe:]CORRection:TRANsducer<1...2>:UNIT?
This command queries the unit (DB|DBVM|DBAM) of the current transducer factor.
The numeric suffix at TRANsducer specifies the primary or secondary transducer.
Example:
CORR:TRAN2:UNIT?
-> DBVM
Return unit (DBVM) of the secondary transducer.
Usage:
Query only
UNIT<1...2>:POWer
<Unit>
This command selects the unit of the vertical axis.
The availability of units depends on the operating mode and type of measurement.
The numeric suffix at UNIT is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<Unit>
DBM | DBMV | DBUV | DUVE | DMVE | VEMF | DBUA | VOLT |
WATT | DUVM | DUAM | V | W | V_M | W_M2 | DB | RHO |
MRHO
Note that the availability of units depends on the operating
mode.
*RST:
DBM
Example:
UNIT:POW DBM
Sets the power unit to dBm.
13.6.6.3
Setting the Bandwidth
The following commands configure the filter bandwidths of the R&S Spectrum Rider.
Note that both groups of commands (BANDwidth and BWIDth) are the same.
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH