![Rohde & Schwarz 1321.0615.02 User Manual Download Page 439](http://html1.mh-extra.com/html/rohde-and-schwarz/1321-0615-02/1321-0615-02_user-manual_1477771439.webp)
Remote Commands
R&S
®
Spectrum Rider FPH
432
User Manual 1321.1011.02 ─ 11
Parameters:
<DisplayRange>
Range:
1 dB to 150 dB
*RST:
100 dB
Default unit: dB
Example:
DISP:TRAC:Y 110dB
Sets the display range to 110 dB.
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]:RLEVel
<RefLevel>
This command defines the reference level.
With a reference level offset ≠ 0, the value range of the reference level is modified by
the offset. You can set the offset with
TRACe<1...2>:Y[:SCALe]:RLEVel:OFFSet
The numeric suffix at DISPlay and TRACe are irrelevant for this command.
Parameters:
<RefLevel>
Sets the reference level; the unit depends on
Range:
Specified in the data sheet.
*RST:
–20 dBm
Default unit: dBm
Example:
DISP:TRAC:Y:RLEV –60dBm
Sets the reference level to -60 dBm.
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]:RLEVel:OFFSet
<RefLvlOffset>
This command defines a reference level offset.
The numeric suffix at DISPlay and TRACe are irrelevant for this command.
Parameters:
<RefLvlOffset>
Sets the reference level offset.
Range:
-100 dB to 100 dB
*RST:
0 dB
Default unit: dB
Example:
DISP:TRAC:Y:RLEV:OFFS –10dB
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]:RPOSition
<RefLvlPosition>
This command defines the position of the reference level on the display grid.
First, you have to set a logarithmic scale for the vertical axis with
WINDow]:TRACe<1...2>:Y:SPACing
on page 319. For a linear scale, you cannot
modify the reference position.
The numeric suffix at DISPlay and TRACe are irrelevant for this command.
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH