![Rohde & Schwarz 1321.0615.02 User Manual Download Page 391](http://html1.mh-extra.com/html/rohde-and-schwarz/1321-0615-02/1321-0615-02_user-manual_1477771391.webp)
Remote Commands
R&S
®
Spectrum Rider FPH
384
User Manual 1321.1011.02 ─ 11
Example:
CALCulate<1...2>:MARKer<1...6>:FUNCtion:
Usage:
Query only
CALCulate<1...2>:MARKer<1...6>:FUNCtion:HARMonics:NHARmonics
<NofHarmonics>
This command sets the number of harmonics. The valid range is from 1 to 6 harmon-
ics.
The suffix at CALCulate and MARKer are irrelevant for this command.
Parameters:
<NofHarmonics>
Range:
1 to 6
*RST:
2
Example:
CALCulate<1...2>:MARKer<1...6>:FUNCtion:
CALCulate<1...2>:MARKer<1...6>:FUNCtion:HARMonics:PRESet
This command optimizes the settings for the harmonic distortion measurement.
This command is an event and therefore has no query and no *RST value.
The suffix at CALCulate and MARKer are irrelevant for this command.
Example:
CALC:MARK:FUNC:HARM ON
Activates the harmonic distortion measurement.
CALC:MARK:FUNC:HARM:NHAR 4
Sets the number of harmonics to 4.
CALC:MARK:FUNC:HARM:PRES
Adjusts the settings of the measurement.
CALC:MARK:FUNC:HARM:LIST?
CALC:MARK:FUNC:HARM:DIST? TOT
Queries the position of the harmonics and the total harmonic dis-
tortion.
Usage:
Event
Measuring the AM Modulation Depth
The following commands configure AM Modulation Depth measurements. To perform
the actual measurement, use the commands described in section
"Performing and Triggering Measurements"
List of commands
CALCulate<1...2>:MARKer<1...6>:FUNCtion:MDEPth[:STATe]
........................................... 385
CALCulate<1...2>:MARKer<1...6>:FUNCtion:MDEPth:RESult?
.......................................... 385