If you set the suffix at DELTamarker to 1, or use no suffix, the R&S Spectrum Rider
interprets this as delta marker 2 because the first marker has to be a normal marker. If
more than one normal marker (2 to 6) is already active, the command turns these
markers into delta markers. If no delta marker is active yet, the command activates the
delta marker and positions it on the trace maximum.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
DELTamarker selects the deltamarker.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:DELT3 ON
Turns on delta marker 3 or turn marker 3 into a delta marker.
CALCulate<1...2>:DELTamarker<1...6>:AOFF
This command turns off all active delta markers.
This command is an event and therefore has no query and no *RST value.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
DELTamarker selects the deltamarker.
Example:
CALC:DELT:AOFF
Turns off all delta markers.
Usage:
Event
CALCulate<1...2>:DELTamarker<1...6>:MAXimum[:PEAK]
This command positions a delta marker on the current trace maximum.
If necessary, the corresponding delta marker is activated first.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
DELTamarker selects the deltamarker.
Example:
CALC:DELT3:MAX
Moves delta marker 3 to the maximum peak.
Usage:
Event
CALCulate<1...2>:DELTamarker<1...6>:MAXimum:NEXT
This command positions a delta marker on the next smaller trace maximum.
If necessary, the corresponding delta marker is activated first.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at