Spectrum Analyzer Mode
R&S
®
Spectrum Rider FPH
153
User Manual 1321.1011.02 ─ 11
line moves down one position, so that the present frame is always on top of the dia-
gram.
Therefore, the sequence of frames is chronological.
1. Press [MEAS] key.
2. Select "Meas Mode" softkey.
The R&S Spectrum Rider opens a submenu to select the measurement mode.
3. Select the "Spectrogram" menu item.
The R&S Spectrum Rider starts the spectrogram result display.
By default, the spectrogram result display consists of two windows. The upper window
shows the measured spectrum as a trace line. The lower window shows the measure-
ment results in a spectrogram. The chronological information in the spectrogram is
restricted by the internal memory of the R&S Spectrum Rider. The R&S Spectrum
Rider stores 1024 frames or spectrums that have been measured in its memory. As the
height of the display is smaller, some of the data becomes invisible after a time.
Figure 6-8: Screen layout for the spectrogram
1 = Operating mode
2 =
3 =
4 = Spectrum result display (optional)
5 = Marker / delta marker (vertical lines)
6 = Time lines T1 and T2 (horizontal lines)
7 = Spectrogram
8 = Scroll direction
9 =
10 = Spectrogram softkey menu