![Rohde & Schwarz 1321.0615.02 User Manual Download Page 145](http://html1.mh-extra.com/html/rohde-and-schwarz/1321-0615-02/1321-0615-02_user-manual_1477771145.webp)
Spectrum Analyzer Mode
R&S
®
Spectrum Rider FPH
138
User Manual 1321.1011.02 ─ 11
6.1.5.1
Selecting a Standard
For more information on selecting a standard, see
Chapter 6.1.3.1, "Selecting a Stan-
6.1.5.2
Optimizing Measurement Settings
After selecting the standard and applying the signal to the R&S Spectrum Rider, you
can optimize measurement settings to avoid overloading the R&S Spectrum Rider.
► Select the "Level Adjust" softkey.
The R&S Spectrum Rider performs a measurement to determine the ideal refer-
ence level and span.
6.1.5.3
Viewing the Results in a Table
You can add a table to the display that shows the measurement results in numerical
form.
1. Press [MEAS] key.
2. Select "View List " softkey.
The R&S Spectrum Rider displays the
Note: You can select "View List" to hide or unhide the
If the list contains more than four entries, you can pan through the list or use the
rotary to navigate the other results. Note that scrolling works only if there is no
active input field.
Measurement items
Descriptions
Tx Power
Power level of the transmission channel.
Tx Bandwidth
Bandwidth of the transmission channel.
PASS / FAIL information
If the signal is within the limits of the spectral
mask, the R&S Spectrum Rider shows
, if
not, it shows
.
Range [Hz]
Frequency range. The first number is the start fre-
quency, the second number the stop frequency of
each defined frequency range. The character fol-
lowing the number indicates the unit (k = kHz, M =
MHz, G = GHz)
RBW [Hz]
Resolution bandwidth the corresponding fre-
quency range is measured with.
Peak
Number of the peak marker (Px).
Performing Spectrum Measurements
www.allice.de