Remote Commands
R&S
®
Spectrum Rider FPH
396
User Manual 1321.1011.02 ─ 11
CALCulate<1...2>:ADEModulation:LIMit:DELete
This command clears the selected limit.
The numeric suffix at CALCulate is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Example:
CALC:ADEM:LIM:DEL
Clears off the selected limit line.
Usage:
Event
CALCulate<1...2>:ADEModulation:LIMit:FAIL?
This command queries the result of a limit check.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end before reading out the result. This is only possible in single sweep mode.
The numeric suffix at CALCulate is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Example:
CALC:ADEM:LIM:FAIL?
Queries the result of the check for limit line
Usage:
Query only
CALCulate<1...2>:LIMit<1...2>:BEEP[:STATe]
<State>
This command turns the beeper that beeps if a limit line is violated on and off.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
LIMit selects the limit line.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:LIM:BEEP ON
Activates the audio beep.
13.6.3.7
Setting the Measurement Mode
The following chapter describes command necessary to switch to the correct measure-
ment mode.
List of commands
Remote Control - Commands