Remote Commands
R&S
®
Spectrum Rider FPH
436
User Manual 1321.1011.02 ─ 11
List of Commands
.........................................................................................436
[SENSe:]BANDwidth[:RESolution]
................................................................................... 436
[SENSe:]BANDwidth[:RESolution]:CISPr
..........................................................................436
[SENSe:]BANDwidth[:RESolution]:CISPr:AUTO
................................................................ 436
[SENSe:]BWIDth[:RESolution]
<ResolutionBW>
[SENSe:]BANDwidth[:RESolution]
<ResolutionBW>
This command defines the resolution bandwidth.
Parameters:
<ResolutionBW>
Range:
1 Hz to 3 MHz
*RST:
- (AUTO is set to ON)
Default unit: GHz
Example:
BAND 100 kHz
Sets the resolution bandwidth to 100 kHz.
[SENSe:]BANDwidth[:RESolution]:CISPr
<CisprResolutionBW>
This command defines the CISPR (6 dB) bandwidth for CISPR conform measure-
ments.
Parameters:
<CisprResolutionBW>200 Hz | 9 kHz | 120 kHz | 1 MHz
*RST:
(AUTO is on)
Example:
BAND:CISP 120KHZ
Defines a bandwidth of 120 kHz.
[SENSe:]BANDwidth[:RESolution]:CISPr:AUTO
<State>
This command turns automatic selection of the CISPR measurement bandwidth on
and off.
Parameters:
<State>
ON | OFF
*RST:
ON
Example:
BAND:CISP:AUTO ON
Turns automatic selection of the CISPR bandwidth on.
13.6.6.4
Performing and Triggering the Measurement
The following commands configure the sweep.
Performing the Measurement
The following chapter describes commands necessary to perform receiver mode mea-
surement.
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH