![Rohde & Schwarz 1321.0615.02 User Manual Download Page 374](http://html1.mh-extra.com/html/rohde-and-schwarz/1321-0615-02/1321-0615-02_user-manual_1477771374.webp)
Remote Commands
R&S
®
Spectrum Rider FPH
367
User Manual 1321.1011.02 ─ 11
EIRP Measurement
This chapter describes all commands available to make an EIRP measurement.
List of commands
........................................................................................... 367
........................................................................................367
...................................................................................... 367
..........................................................................................367
......................................................................................... 368
.......................................................................................... 368
.............................................................................................368
DISPlay<1...2>:EIRP:HORizontal
.....................................................................................368
DISPlay<1...2>:EIRP:MAXHold:RESet
............................................................................. 368
........................................................................................369
[SENSe:]EIRPpower:CLOSs
<EIRPCableLoss>
This command input cable loss value in dB and value is displayed on the EIRP info
bar.
Parameters:
<EIRPCableLoss>
Numeric value that defines the cable loss.
Default unit: dB
Example:
EIRP:CLOS 0
Cable loss of 0.00 dB is displayed on the EIRP info bar.
[SENSe:]EIRPpower:DISTance
<EIRPDistance>
This command input parameter distance in meter or feet and value is displayed on the
EIRP info bar.
Parameters:
<EIRPDistance>
Numeric value that defines the distance.
Default unit: m
Example:
EIRP:DIST 0.95
Distance of 950 mm is displayed on the EIRP info bar.
[SENSe:]EIRPpower:MAXHold?
This command queries the maximum hold value of the EIRP measurements.
Usage:
Query only
[SENSe:]EIRPpower:PLOSs?
This command queries the free space path loss calculation.
FSPL
dB
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH