Remote Commands
R&S
®
Spectrum Rider FPH
425
User Manual 1321.1011.02 ─ 11
Parameters:
<Unit>
DBM | DB | WATT | VSWR | W | DUVM | DBUV | DBMV |
DUVE | DMVE | W_M2 | V_M
Note on the parameter DB: when applied to UNIT1, the power is
relative to the reference level, when applied to UNIT2, the return
loss is displayed.
Note on the parameter VSWR: the parameter is only available if
applied to UNIT2.
Example:
UNIT1:PMET:POW DBM
When measuring with the R&S FSH-Z1, R&S FSH-Z18 or USB
power sensors: sets unit to dBm.
When measuring with the R&S FSH-Z14 or R&S FSH-Z44: sets
unit of forward power to dBm.
13.6.5.4
Defining the Measurement Time
The following chapter describes commands to define the measurement time of the
power sensor.
List of commands
................................................................................................ 425
[SENSe:]PMETer:MTIMe
<MeasTime>
This command sets the duration of measurements.
Available for measurements with a power sensor.
Parameters:
<MeasTime>
SHORt | NORMal | LONG
Example:
PMET:MTIMe SHOR
Sets a short measurement time for power measurements.
13.6.5.5
Zeroing of the Power Sensor
The following chapter describes commands to perform zeroing of the power sensor.
List of commands
CALibration<1...2>:PMETer:ZERO:AUTO
......................................................................... 425
CALibration<1...2>:PMETer:ZERO:AUTO
<ONCE>
This command starts to zero the power sensor.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALibration is irrelevant for this command.
Remote Control - Commands