Instrument Functions
R&S
®
Spectrum Rider FPH
98
User Manual 1321.1011.02 ─ 11
4.3
Presetting the R&S Spectrum Rider
Before you prepare a measurement, it is recommended to preset the R&S Spectrum
Rider. During a preset, the R&S Spectrum Rider resets all settings to their default
state. Restoring the default configuration has the advantage that old settings do not
affect measurements.
The default setup is specific to the operating mode.
► Press [PRESET] key.
The R&S Spectrum Rider restores its default setup.
You can also define your own default settings via a dataset. This dataset is then loaded
after pressing the [PRESET] key instead of the factory default.
1. Press [SETUP] key.
2. Select "User Preference" softkey.
3. In the "User Preferences" dialog box, select the "Preset Dataset" item.
The R&S Spectrum Rider displays the file manager dialog box to select the dataset
that contains the settings you would like to have as the preset settings.
4. Select the dataset with the settings you want and press "Load" softkey.
5. In the "User Preferences" dialog box, select the "Preset Mode" item.
A drop-down menu opens to select the preset mode.
6. Select "User Defined" to load the dataset defined in the "Preset Dataset".
The R&S Spectrum Rider now loads the settings of the dataset after you press
[PRESET.]
Presetting the R&S Spectrum Rider